ǥ :
|
ȣ - 550051 190 |
Effect of Atrial Septal Defect Shape Evaluated Using 3-Dimensional Transesophageal Echocardiography on Size Measurements for Percutaneous Closure |
인제대학교 부산백병원¹ 울산대학교 서울아산병원² |
서정숙¹ , 송종민² , 김영학² , 박덕우² , 이승환² , 김원장² , 김대희² , 강덕현² , 송재관² |
Background: The effect of ASD shape on ASD size measurements remains unclear. This study was performed to assess the relationships between stretched balloon diameter (SBD) and diameters measured by 2-dimensional (2D) and 3-dimensional (3D) transesophageal echocardiography (TEE), and to evaluate the effect of atrial septal defect (ASD) shape on these relationships in patients with secundum ASD.
Methods: We prospectively enrolled 107 consecutive patients who underwent transcatheter closure of ASD. SBDs and ASD diameters on 2D and 3D TEE were measured. The circular index of ASD was defined as the ratio of maximal to minimal diameter on 3D TEE image. All patients were clinically followed up for a median of 1.3 months after the procedure.
Results: The correlations between measurements were significantly better in patients with round (circular index < 1.5) than with ovoid ASD (Figure). The differences between SBDs and maximal diameters measured by 3D TEE were significantly smaller in patients with large ovoid ASDs than in other patients (Table). The differences between the size of finally implanted devices and the maximal diameters measured by 3D TEE were smaller in patients with large ovoid than with round ASDs. There were no major complications, including device embolization, during follow-up.
Conclusions: The relationships between SBD and maximal diameters measured by 2D and 3D TEE are influenced by ASD shape. All parameters including ASD shape should be considered in selecting device size for transcatheter ASD closure.
|
|
Round-small (N=38) |
Round-large (N=27) |
Ovoid-small (N=26) |
Ovoid-large (N=16) |
P-value |
3Dmax-2Dmax |
-0.9±1.2* |
-1.0±1.5* |
-0.2±2.0 |
0.6±2.5 |
0.006 |
SBDfluoro-3Dmax |
2.0±1.7* |
1.6±2.2* |
2.4±2.8* |
-0.5±3.5 |
0.002 |
Device-SBDfluoro |
1.1±1.6† |
1.9±2.0† |
-0.2±1.8* |
1.5±2.5 |
0.001 |
Device-3Dmax |
3.1±1.7* |
3.5±1.6* |
2.2±2.2 |
1.1±2.3 |
<0.001 |
2Dmax = maximal diameter on 2-dimensional transesophageal echocardiography (TEE), 3Dmax = maximal diameter on 3-dimensional TEE, SBDfluoro = stretched balloon diameter on fluoroscopy.
* p<0.05 vs. Ovoid-large, † p<0.05 vs Ovoid-small
|
Warning: getimagesize(/home/virtual/circulationadmin/renewal/econgress/conference/abstract/img_files/그림1-resize.jpg) [function.getimagesize]: failed to open stream: No such file or directory in /home/virtual/circulationadmin/new/econgress/conference/manage/schedule/view_abstract.php on line 164
|
|